Reverberation rooms
Two rooms, used two ways. Singly for absorption to ISO 354 and ASTM C423. As a coupled pair for sound insulation to ASTM E90 and ISO 10140-2, testing building elements at the size they are actually installed at.
Diffusion is the whole trick
A reverberation room is not simply a hard room. The method assumes a diffuse field — sound energy arriving at the specimen equally from every direction — and everything depends on that assumption holding. Parallel walls produce standing waves that concentrate energy at particular frequencies and particular positions, which is exactly what a diffuse field must not have. Suspended diffusers break those patterns up, and the room is then qualified by test to confirm the field is diffuse enough over the frequency range that gets reported.
This is why a measured absorption coefficient occasionally exceeds 1.0, which sounds like an error and is not. The coefficient is referenced to the specimen area, and edge effects mean a finite specimen absorbs more than its own area suggests. It is a known and documented property of the method, not a laboratory mistake — and a supplier who quietly truncates it to 1.00 is hiding the fact that they know it.
Before you ship anything
Send us the element dimensions and the build-up first. Two things get decided at that point and both are expensive to discover late: whether your element fits the test opening, and whether it can physically be brought into the room already assembled or has to be built up here. A pre-assembled unit that cannot pass the test room door is a wasted delivery, and it happens.
What we can test
- Specimens accepted
- Doors, windows, glazing units, facade panels and partition assemblies, tested at the size they are installed at. Send your element dimensions and we confirm the fit before you ship
- Absorption specimen area
- 10–12 m², per ISO 354
- Mounting conditions available
- Type A, Type E-200, Type J
- Valid frequency range
- 100–5,000 Hz, one-third-octave bands
- Room qualification
- Fully qualified to ISO 354 Annex A for absorption testing
- Diffusion
- Suspended panel diffusers, qualified per ISO 354
Full room data is issued with your quotation
Room volumes, background noise levels and the full qualification record are commercially sensitive and are provided with your quotation and on every test report, as ISO/IEC 17025 requires. What we publish here is what you need in order to decide whether we can test your product.