Sound level meter calibration to IEC 61672-3 — ISO/IEC 17025 accredited
Sound level meter calibration at ALT is periodic testing to IEC 61672-3, verifying conformity to the IEC 61672-1 specification for Class 1 and Class 2 instruments. It is inside our ISO/IEC 17025 accredited scope, certificate 24-LB0054. We calibrate every make: the meter does not have to have been bought from us, because we do not sell meters.
What a sound level meter calibration covers
Periodic testing under IEC 61672-3 is a sequence of acoustical and electrical tests that together establish whether the instrument still meets the tolerance limits of its declared class. It is not a single check at one level and one frequency. Each result is reported against the limit, with the measurement uncertainty, so a reviewer can see the margin rather than only a pass mark.
- Acoustic signal tests at the calibration check frequency
- Self-generated noise
- Frequency weightings A, C and Z
- Time weightings F, S and I
- Level linearity across the operating range
- Toneburst response
- Overload indication
- Long-term stability against the previous certificate where available
Included at no extra charge
Incoming condition check, battery and power check, cleaning, fault diagnosis and a firmware check. If we find a fault we quote the repair before doing anything, and recalibrate after repair.
How often should a sound level meter be calibrated?
Once a year is what we recommend, and it is what most accreditation bodies, project specifications and occupational noise programmes expect to see on the certificate. An annual interval also means the long-term stability test has something meaningful to compare against, which is how drift is caught before it costs you a disputed measurement.
Shorten it if the instrument is used daily, travels to site constantly, works in heat, dust or humidity, has been dropped, or produces numbers that go into a contract or a legal case. Lengthening it is a decision for the owner's quality system to justify, not a default.
A periodic calibration does not replace the field check you do with an acoustic calibrator before and after each measurement session. The field check confirms the instrument has not shifted since yesterday; the calibration establishes that it conforms to its class across the whole range.
Accredited parameters and measurement capability
The table below is taken from our accredited scope under certificate 24-LB0054. The calibration and measurement capability is the smallest uncertainty we can report for that parameter, at approximately 95% confidence. A laboratory that cannot show you this table for its own scope is not offering the same service.
| Parameter | CMC (±) |
|---|---|
| Indication at the calibration check frequency | 0.12 dB |
| Self-generated noise | 0.060 dB |
| Electrical signal test of frequency weightings | 0.13 dB |
| Frequency and time weighting at 1 kHz | 0.13 dB |
| Long-term stability | 0.10 dB |
| Level linearity on the reference level range | 0.14 dB |
| Level linearity including the level range control | 0.14 dB |
| Tone burst response | 0.14 dB |
| Peak C sound level | 0.13 dB |
| Overload indication | 0.13 dB |
| High level stability | 0.10 dB |
What you receive
- An ISO/IEC 17025 accredited calibration certificate carrying accreditation number 24-LB0054, issued in English.
- Measured values against the IEC 61672-1 tolerance limits for the declared class, with the measurement uncertainty for each.
- A clear conformity statement, and where the instrument fails, which parameter failed and by how much.
- The calibration date and the instrument identification as it will appear in your own equipment register.
- A statement of traceability to national standards.
Sound level meters and analysers we calibrate
- Brüel & Kjær (HBK) — including the 2250, 2270 and 2245 families
- Norsonic — Nor131, Nor140, Nor145, Nor150 and earlier models
- Rion — NL-42, NL-52, NL-62 and the NA analyser series
- Svantek — SVAN 971, 977, 979 and related models
- 01dB (Acoem) — Fusion, Cube, Solo
- Larson Davis (PCB Piezotronics) — 831, 831C, SoundTrack LxT
- Casella, Cirrus Research, NTi Audio, Pulsar Instruments, Ono Sokki, Sinus Messtechnik and Wölfel
- General-purpose and Class 2 meters from Extech, PCE, Testo, Delta OHM and similar makers
We are not an authorised service centre for any of these makers and have no commercial relationship with them. Calibration is performed against the standard, not against a manufacturer's procedure, so the make matters far less than the class of the instrument. If yours is not listed, ask — the list is what we see most, not a limit.
Turnaround and how to book
Service
- Method
- IEC 61672-3
- Classes
- Class 1, Class 2
- Standard lead time
- 10 working days from receipt
- Express service
- Available on request, subject to laboratory schedule — quoted at booking
- Loan instrument during calibration
- On request, subject to availability
- How to start
- Send the calibration requisition, or contact us by email, telephone or LINE @altbkk
Common questions about sound level meter calibration
How often should a sound level meter be calibrated?
Once a year is the interval we recommend, and the one most accreditation bodies, project specifications and occupational noise programmes expect. Shorten it for daily use, site work, harsh conditions, a dropped instrument, or measurements that go into a contract or a legal case. This is separate from the field check you do with a calibrator before and after each session.
What is IEC 61672-3?
IEC 61672-3 is the periodic testing standard for sound level meters. It sets out the tests a laboratory performs to establish whether an instrument still conforms to the IEC 61672-1 specification for its declared class. IEC 61672-1 is the specification itself; IEC 61672-3 is how conformity is checked after the instrument is in service.
What is the difference between a field check and a calibration?
A field check uses an acoustic calibrator to confirm the instrument reads correctly at one level and one frequency, usually 94 dB at 1 kHz. It takes seconds and should be done before and after every measurement. A periodic calibration to IEC 61672-3 tests weightings, linearity across the range, self-noise, toneburst response, overload and stability, and produces an accredited certificate. One is a daily sanity check, the other is evidence.
Do you calibrate meters that were not bought from you?
Yes, and that is the normal case. ALT does not sell, distribute or represent any instrument brand, so there is no make we prefer and no make we decline on commercial grounds. We calibrate against the standard rather than against a manufacturer procedure, so the class of the instrument matters far more than the badge on it.
Does the microphone need to be calibrated separately?
For periodic testing to IEC 61672-3 the meter is tested as a system, with its own microphone fitted. A separate microphone calibration to IEC 61094-2 or IEC 61094-6 is a different accredited service, and is what you need when the microphone is used on more than one instrument, when free-field or pressure response data is required in its own right, or when a specification asks for the microphone certificate.
What happens if my sound level meter fails the calibration?
We tell you which parameter failed and by how much, before anything else happens. Nothing is adjusted or repaired without a quotation you have accepted. If a repair goes ahead, the instrument is recalibrated afterwards and the certificate reflects the state it leaves in. If it is not economic to repair, we say so plainly and do not recommend a supplier.
How long does calibration take, and is it accredited?
Ten working days from receipt is the standard lead time, with an express route available on request. Sound level meter calibration to IEC 61672-3 is inside ALT's ISO/IEC 17025 accredited scope under certificate 24-LB0054, alongside sound calibrators and measurement microphones.